FDA Medical Device Recalls (openFDA)
Insulet Corporation: The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.
- Recall number
- Z-0132-2026
- Recalling firm
- Insulet Corporation
- Product
- The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.
- Status
- Open, Classified
- Initiated
- 2025-09-11
- Posted
- 2025-10-10
- Terminated
- not on file
- Reason
- The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
- Root cause
- Software design
- Action
- Insulet notified both healthcare professionals and users on about 09/11/2025 via URGENT: Medical Device Correction letter. Instructions included that an update to the application will be pushed to all devices and should be implemented by users. Users were also requested to confirm receipt and acknowledgement of the notification but selecting the "acknowledgement button" in the email.
- Quantity
- 174,013 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X
- 510(k) numbers
- ["K231826"]
- PMA numbers
- not on file
- Product code
- QRX
- Firm FEI
- 3014585508
- Firm city
- Acton
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28