Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: CORE Suction Irrigation Handpiece without Probe for Single Solution Bags, Catalog Number CD8300 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures.

Recall number
Z-0132-2017
Recalling firm
ConMed Corporation
Product
CORE Suction Irrigation Handpiece without Probe for Single Solution Bags, Catalog Number CD8300 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures.
Status
Terminated
Initiated
2016-06-27
Posted
not on file
Terminated
2018-06-07
Reason
The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
Root cause
Under Investigation by firm
Action
ConMed sent an Urgent Device Recall letter dated June 27, 2016, to all affected customers with reply forms via USP First Class Mail. International customers received a notice dated July 11, 2016 by courier service. Customers were advised to inspect their inventory for any of the devices subject to recall. Customers were urged to contact all departments within their facility and any other facilities within their organization that may have received affected products. It is imperative that all end users of these devices receive the notification. Customers who have received any of the affected product are asked to complete and return the reply form along with the device to: CONMED Corporation 525 French Road Utica, NY 13502 USA Attn: Ed Kovac. Please do not return used devices. Replacement product will be issued unless otherwise requested. If customers do not have any affected devices to return, complete and return the business reply form indicating they have no devices and return by one of the means listed:1. Email to: coresihp@conmed.com 2. Fax to: Field Action Support Team at 315-624-3225. 3. US mail, using enclosed postage-paid envelope, to: CONMED, 525 French Road, Utica, NY 13502; Attn: Field Action Support Team. Customers with questions or requests, should contact the Field Action Support Team at 315-624-3237, fax to 315-624-3225, or email coresihp@conmed.com.
Quantity
not on file
Distribution
Worldwide Distribution - US (Nationwide) Internationally: to Australia, Austria, Uruguay, Denmark, Germany, Ireland, Italy, United Kingdom, Guatemala, Czech Republic & Canada.
Lot, serial or model codes
Beginning Lot Code 110609X to Ending Lot Code 20160308X
510(k) numbers
["K926477"]
PMA numbers
not on file
Product code
GBW
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register