Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories, Inc: ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT System is designed to perform automated chemistry tests utilizing photometry and potentiometric technology; and immunoassay tests utilizing chemiluminescent microparticle assay detection technology.

Recall number
Z-0132-2010
Recalling firm
Abbott Laboratories, Inc
Product
ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT System is designed to perform automated chemistry tests utilizing photometry and potentiometric technology; and immunoassay tests utilizing chemiluminescent microparticle assay detection technology.
Status
Terminated
Initiated
2009-07-17
Posted
2009-11-09
Terminated
2010-01-26
Reason
Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.
Root cause
Other
Action
A product correction notice was issued to consignees on July 17, 2009. Field representatives were instructed to visit consignees, inspect cables and replace them
Quantity
4,526
Distribution
Worldwide Distribution
Lot, serial or model codes
ARCHITECT i2000 System List Number 8C89-01; ARCHITECT i2000SR System System List Number 3M74-01 and 3M74-02; Wash Zone Mechanism Grounding Kit Part Number 7-201369-01; Concentrated Wash Buffer 1 L List Number 6C54-58 and l0 L List number 6C54-88.
510(k) numbers
["K983212"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register