Skip to the content

FDA Medical Device Recalls (openFDA)

Contego Medical LLC: Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40

Recall number
Z-0131-2026
Recalling firm
Contego Medical LLC
Product
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40
Status
Open, Classified
Initiated
2025-09-04
Posted
2025-10-10
Terminated
not on file
Reason
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.
Root cause
Device Design
Action
On 9/4/2025, communication and training to the field representatives started with in person letter distribution following immediately after. Customers were asked to do the following: Action Required 1) Inventory Check: Medtronic and Contego sales representatives will visit your facility to conduct an inventory count of devices from the affected lots. 2) Quarantine Affected Devices: Any identified devices will be removed from use and quarantined by the Medtronic or Contego team. 3) Return: Medtronic and Contego will remove affected devices. If replacements are desired, they can be provided during the same visit. 4) Confirmation: Sign and return the enclosed acknowledgment form to confirm receipt of this advisory and the completion of actions performed by Medtronic or Contego. 5) Please email the completed form to Contego Medical recall@contegomedical.com as soon as possible and retain a copy of the form to provide to the representative at the time of their visit. If you have questions, contact the firm's Recall Support Team at recall@contegomedical.com or call 919-459-7250 Extension 4.
Quantity
568
Distribution
US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): NG-0740-140-2/00851616007613/Z2460707D(Z2460707D), Z2460733A(03/05/2027), Z2460766A(04/07/2027), Z2460774B(04/17/2027), Z2460780A(04/18/2027); NG-0640-140-2/00851616007590/Z2460746C(03/20/2027), Z2460774A(04/18/2027); NG-NV-7-40/00851616007620/Z2460776D(04/14/2027); NG-NV-6-40/00851616007651/Z2460776B(04/16/2027)
510(k) numbers
not on file
PMA numbers
["P240009"]
Product code
NIM
Firm FEI
3011471056
Firm city
Raleigh
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register