Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance - Product Usage: The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.

Recall number
Z-0131-2021
Recalling firm
GE Healthcare, LLC
Product
Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance - Product Usage: The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.
Status
Terminated
Initiated
2020-09-08
Posted
not on file
Terminated
2021-04-28
Reason
Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
Root cause
Process change control
Action
On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
Quantity
not on file
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
Lot, serial or model codes
Model: 5726207-8 and 5741000-28
510(k) numbers
["K163619"]
PMA numbers
not on file
Product code
OUO
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register