FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015
- Recall number
- Z-0130-06
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015
- Status
- Terminated
- Initiated
- 2005-02-10
- Posted
- 2005-11-16
- Terminated
- 2008-11-03
- Reason
- Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.
- Root cause
- Other
- Action
- Baxter mailed Urgent Device Correction letters dated 2/10/05 to the domestic accounts on 2/10/05 via first class mail, to the attention of the Director of Nursing, advising them of the following hardware problems and software anomalies and providing workarounds for each. There are potential failures of the pump due to upstream occlusion alarms, electro-static discharge, Key-Only security mode (previously covered by letter dated 4/10/02), system error 32 or 33 codes, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. The accounts were advised that as soon as revised software and hardware is available, Baxter would modify the pumps at no additional charge to the consignee. Any questions were directed to their Baxter Medication Delivery Product Information Center at 1-800-933-0303.
- Quantity
- 24,009 pumps
- Distribution
- Nationwide including Puerto Rico, and internationally through Baxter subsidiaries in Canada, Chile, Hong Kong, Kuwait and Saudi Arabia.
- Lot, serial or model codes
- product codes 2L3107 and 2L3107R, all serial numbers
- 510(k) numbers
- ["K993387"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28