Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015

Recall number
Z-0130-06
Recalling firm
Baxter Healthcare Corp.
Product
Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL 60015
Status
Terminated
Initiated
2005-02-10
Posted
2005-11-16
Terminated
2008-11-03
Reason
Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.
Root cause
Other
Action
Baxter mailed Urgent Device Correction letters dated 2/10/05 to the domestic accounts on 2/10/05 via first class mail, to the attention of the Director of Nursing, advising them of the following hardware problems and software anomalies and providing workarounds for each. There are potential failures of the pump due to upstream occlusion alarms, electro-static discharge, Key-Only security mode (previously covered by letter dated 4/10/02), system error 32 or 33 codes, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. The accounts were advised that as soon as revised software and hardware is available, Baxter would modify the pumps at no additional charge to the consignee. Any questions were directed to their Baxter Medication Delivery Product Information Center at 1-800-933-0303.
Quantity
24,009 pumps
Distribution
Nationwide including Puerto Rico, and internationally through Baxter subsidiaries in Canada, Chile, Hong Kong, Kuwait and Saudi Arabia.
Lot, serial or model codes
product codes 2L3107 and 2L3107R, all serial numbers
510(k) numbers
["K993387"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register