Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.

Recall number
Z-0127-2014
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.
Status
Terminated
Initiated
2013-11-04
Posted
2013-11-01
Terminated
2014-04-29
Reason
Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m
Root cause
Device Design
Action
On 11/4/13, an URGENT MEDICAL DEVICE Recall Notification was sent to all domestic consignees via mail and foreign by email. An advisory was sent to every physician that uses any of the products listed. Every physician that uses any of these products must print their name and sign below which means you acknowledge that youve received a copy of the attached advisory and that you understand the information regarding the proper usage of the Decompressor. Once the form has been signed by all physicians, the form must be completed and signed by the recall coordinator. Please fax to 866-521-2762 or email to kara.spath@stryker.com.
Quantity
20,385 total distributed since 2004
Distribution
Nationwide Distribution including the states of AR, AZ, CA, CO, FL, GA, IL, KS, MS, ND, and NE.
Lot, serial or model codes
Part Number(s) 0407-251-000, 0407-260-000, 0407-265-000, 0407-266-000, 0407-280-000, and 0407-281-000.
510(k) numbers
["K032473"]
PMA numbers
not on file
Product code
HRX
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register