Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc: Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA

Recall number
Z-0127-2011
Recalling firm
Hospira Inc
Product
Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
Status
Terminated
Initiated
2010-02-22
Posted
2010-10-22
Terminated
2013-12-03
Reason
Hospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
Root cause
Device Design
Action
Hospira issued a Recall Notification letter dated October 4, 2010, updating the Urgent Device Field Correction letter dated February 22, 2010, ideentifying the affected products, the issue prompting the recall, and actions to be taken by customers. Customers are to examine their inventory to determine if they have any of the affected products, and recommended to remove them from service in critical care areas, and contact Hospira to schedule removal and replacement of the motor. Replace affected devices with a loaner or corrected back-up pump if available. Custoemrs can contacvtd Hospira Global Product Safety and Complaints at 1-800-441-4100, and Technical Support Operations at 1-800-241-4002.
Quantity
not on file
Distribution
Worldwide Distribution: USA and the coutnries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
Lot, serial or model codes
All units of these List Numbers.
510(k) numbers
["K041550"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register