FDA Medical Device Recalls (openFDA)
Hospira Inc: Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
- Recall number
- Z-0127-2011
- Recalling firm
- Hospira Inc
- Product
- Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
- Status
- Terminated
- Initiated
- 2010-02-22
- Posted
- 2010-10-22
- Terminated
- 2013-12-03
- Reason
- Hospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
- Root cause
- Device Design
- Action
- Hospira issued a Recall Notification letter dated October 4, 2010, updating the Urgent Device Field Correction letter dated February 22, 2010, ideentifying the affected products, the issue prompting the recall, and actions to be taken by customers. Customers are to examine their inventory to determine if they have any of the affected products, and recommended to remove them from service in critical care areas, and contact Hospira to schedule removal and replacement of the motor. Replace affected devices with a loaner or corrected back-up pump if available. Custoemrs can contacvtd Hospira Global Product Safety and Complaints at 1-800-441-4100, and Technical Support Operations at 1-800-241-4002.
- Quantity
- not on file
- Distribution
- Worldwide Distribution: USA and the coutnries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.
- Lot, serial or model codes
- All units of these List Numbers.
- 510(k) numbers
- ["K041550"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 2921482
- Firm city
- Morgan Hill
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28