AGA Medical Corporation: AMPLATZER Vascular Plug, Manufactured by AGA Medical Corporation , 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. Sterile EO. Single Use Only. Rx Only. Do not use open or damaged packages. Store in a cool, dry place. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device made from a Nitinol wire mesh. The device is secured on both ends with platinum marker bands. A stainless steel micro screw is welded to one of the platinum marker bands, which allows attachment to the 135 cm long deliver cable. The AMPLATZER Vascular Plugs are made in various sizes ranging from 4 to 16 mm in 2 millimeter increments.

Recall number
Z-0127-06
Recalling firm
AGA Medical Corporation
Product
AMPLATZER Vascular Plug, Manufactured by AGA Medical Corporation , 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. Sterile EO. Single Use Only. Rx Only. Do not use open or damaged packages. Store in a cool, dry place. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device made from a Nitinol wire mesh. The device is secured on both ends with platinum marker bands. A stainless steel micro screw is welded to one of the platinum marker bands, which allows attachment to the 135 cm long deliver cable. The AMPLATZER Vascular Plugs are made in various sizes ranging from 4 to 16 mm in 2 millimeter increments.
Status
Terminated
Initiated
2005-09-01
Posted
2005-11-05
Terminated
2006-11-13
Reason
Five lots of AMPLATZER Vascular Plugs were sterilized in a load that was not validated for the AMPLATZER Vascular Plug. However, only three of those five lots were distributed.
Root cause
Other
Action
Urgent Recall letters, dated 09/01/05, were sent to affected customers via email, facsimile and with follow-up delivery via USPS. OUS customers were notified via Federal Express. The letter described the situation, the lots being recalled and included a response form that was to be faxed back to AGA Medical.
Quantity
81
Distribution
OUS countries include: Australia, Germany, Italy, Korea, Malasia, Slovakia, Spain, Switzerland, Thailand, United Kingdom
Lot, serial or model codes
The three lots that were distributed are identified as Lot no. M05F10-11, M05F15-12 and M05F02-47.
510(k) numbers
["K031810"]
PMA numbers
not on file
Product code
KRD
Firm FEI
3005334138
Firm city
Golden Valley
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register