Skip to the content

FDA Medical Device Recalls (openFDA)

SynCardia Systems LLC: SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO

Recall number
Z-0126-2024
Recalling firm
SynCardia Systems LLC
Product
SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
Status
Completed
Initiated
2023-08-01
Posted
2023-10-20
Terminated
not on file
Reason
Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
Root cause
Other
Action
On August 22, 2023, SynCardia issued an Urgent Medical Device Correction Notification to all affected consignees via Email. SynCardia asked consignees to take the following actions: 1. You may continue to use the device while the change undergoes FDA review for safety and effectiveness. 2.Ensure that all medical professionals in your institution are informed of this Urgent Medical Device Correction. 3. Please complete the Customer Acknowledgement Form and return it via email to SynCardia Systems, LLC, within five business days of receipt of this letter. 4.For patients with implanted devices affected by this correction, monitor patients closely and take appropriate actions you believe necessary if any issues arise. 5. Adverse reactions or quality problems experienced with this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download the form from www.fda.gov/medwatch/getforms.htm or call 1-800-322-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Quantity
114 units
Distribution
US: FL, WI, AZ, TX, CA, OH, NC, IN, OK, IL, VA, NY, NE, PA, WA, OUS: Saudi Arabia, Austria, Italy, Macedonia, Turkey, Germany, France, Kuwait, Serbia, United Kingdom, Canada, Lithuania, Slovenia Australia
Lot, serial or model codes
All Lots/ UDI: 00858000003053
510(k) numbers
not on file
PMA numbers
["P030011"]
Product code
LOZ
Firm FEI
3003761017
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register