Skip to the content

FDA Medical Device Recalls (openFDA)

Carestream Health, Inc.: Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.

Recall number
Z-0126-2014
Recalling firm
Carestream Health, Inc.
Product
Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.
Status
Terminated
Initiated
2013-06-06
Posted
2013-10-31
Terminated
2016-05-20
Reason
Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.
Root cause
Device Design
Action
On June 6, 2013 Carestream Health Inc. initiated a recall for the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch. Carestream Health Inc. will replace the power plug in all potentially impacted DRX-Revolution Mobile X-Ray Systems. Your Carestream service representative will contact you to answer any questions and schedule a visit to replace the plug of your Mobile X-Ray System as soon as possible. If you have any questions or concerns, or if you experience any issue with your plug prior to our visit, please contact the Carestream Customer Care Center at 1-800-328-2910, 8AM through 8pm Eastern Standard Time.
Quantity
180 units
Distribution
Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.
Lot, serial or model codes
Serial Numbers: 111, 114, 119-123, 125-126, 134, 142-143, 151-155, 161-169, 173-178,180-183, 188-192, 196-209, 211, 221-232, 235-249, 254, 261-270, 275-276, 279-286, 305, 312, 316-326, 330-337, 339-349, 353, 355, 357-365, 372-373, 375-376, 379-385, 387-388, 394-401, 403-405, 409-410, 412-415, 417, 420-421, 423, 431-435, 441-446, 451-461, 464-470, and 482-490
510(k) numbers
["K120062"]
PMA numbers
not on file
Product code
IZL
Firm FEI
1315356
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register