Skip to the content

FDA Medical Device Recalls (openFDA)

Nicolet Biomedical Div of Viasys Healthcare: Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711

Recall number
Z-0126-2008
Recalling firm
Nicolet Biomedical Div of Viasys Healthcare
Product
Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711
Status
Terminated
Initiated
2007-09-11
Posted
2008-01-03
Terminated
2011-12-10
Reason
Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.
Root cause
Component design/selection
Action
Certified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.
Quantity
33
Distribution
Worldwide Dsitribution - USA, Germany, Netherlands, Kuwait, Japan, Ireland, and China.
Lot, serial or model codes
ON040060, ON050120, ON050130, ON050168, ON050186, ON050197, ON050199, ON050203, ON050206, ON050212, ON050220, ON050231, ON060232, ON060243, ON060254, ON060256, ON060257, ON060269, ON060270, ON060272, ON060275, ON060276, ON060277, ON060278, ON060279, ON060280, ON060288, ON060290, ON060291, ON060292, ON060299, ON070301, ON070308.
510(k) numbers
["K021185"]
PMA numbers
not on file
Product code
GWQ
Firm FEI
3010611950
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register