Trivascular, Inc: Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).
- Recall number
- Z-0125-2015
- Recalling firm
- Trivascular, Inc
- Product
- Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).
- Status
- Terminated
- Initiated
- 2014-09-20
- Posted
- 2014-11-03
- Terminated
- 2014-12-04
- Reason
- TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
- Root cause
- Under Investigation by firm
- Action
- Urgent Field Safety Notice letter was sent to all consignees on 9/19/14 via trackable mail system. Users are notified that TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.
- Quantity
- 0 Devices
- Distribution
- International Distribution Only -- Canada, Turkey, Austria, Ireland, Belgium, Italy, Switzerland, Spain, France, and Cyprus.
- Lot, serial or model codes
- The devices subject to this action are labeled for distribution outside of the United States; therefore, no UDI is on the device label or package. Model Number: TV-AB2980-D Lot Numbers: FS031814-11 FS052214-22 FS052814-36 FS040414-21 FS050614-56 FS040714-53 FS040514-02 FS042614-05 FS050614-53 FS032614-46 FS050214-27 FS050314-12 FS052214-48 FS052214-53 FS031814-44 FS031814-49 FS032714-04 FS032714-20 FS040414-30 FS040514-08 FS050114-01 FS050114-08 FS050514-18 FS050514-27 FS050614-54 FS050614-55 FS051914-32 FS051914-34 Expiry dates ranging "May-17" through "Jun-17"
- 510(k) numbers
- not on file
- PMA numbers
- ["P120006"]
- Product code
- MIH
- Firm FEI
- 3008011247
- Firm city
- Santa Rosa
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28