Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 A medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems.

Recall number
Z-0125-2013
Recalling firm
GE Healthcare, LLC
Product
GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 A medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems.
Status
Terminated
Initiated
2012-05-24
Posted
2012-10-24
Terminated
2012-12-11
Reason
GE Healthcare has recently become aware of inaccurate Scale factor annotation provided by SenoAdvantage 2.2 workstations that may impact patient safety. This issue may result in a potential miscalculation of size.
Root cause
Software design
Action
Consignees were sent on 5/24/12 a GE Healthcare "Urgent Medical Device Correction" letter dated May 23, 2012. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, and Product Correction. Customers should verify if their workstation is affected. GE Healthcare will correct all affected workstations by providing a software upgrade and Operator Manual update. A GE Healthcare service representative will contact you to arrange this correction. This activity will be performed at no cost to you. If you have any questions or concerns regarding this notification, please call one of the following phone numbers: United States: 800 437 1171 Japan: 0120 - 055 - 919 For other countries, please contact your local GE Healthcare Service Representative.
Quantity
12
Distribution
Worldwide Distribution -- USA (nationwide) including states of: CA, CT, NY, TX, and MS and countries of TAIWAN, ITALY, CANADA, and BAHRAIN
Lot, serial or model codes
000000010005GS 000000010004GS 000000010009GS 000000010007GS 000000010006GS 000000010010GS 000000010001GS 000000010002GS 000000010008GS 000000010003GS 00000L3A4391A7 00000L3A8D5CF0
510(k) numbers
["K033400"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register