Randox Laboratories Ltd.: Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.

Recall number
Z-0123-2017
Recalling firm
Randox Laboratories Ltd.
Product
Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
Status
Terminated
Initiated
2016-08-02
Posted
2016-10-18
Terminated
2017-09-07
Reason
According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.
Root cause
Nonconforming Material/Component
Action
Randex Laboratories sent an Urgent Field Safety Notice letter dated June 17, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the response section of this form technical.services@randox.com. For questions contact Randex Technical Service.
Quantity
18
Distribution
US Nationwide Distribution
Lot, serial or model codes
Catalogue number:   LI3837, LI7979, LI8050, LI8361
510(k) numbers
not on file
PMA numbers
not on file
Product code
CDT
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register