Zimmer, Inc.: The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate.

Recall number
Z-0123-2015
Recalling firm
Zimmer, Inc.
Product
The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the Persona Tibial Articular Inserter is used to seat the articular surface implant onto the tibial baseplate.
Status
Terminated
Initiated
2014-10-07
Posted
2014-10-23
Terminated
2015-09-23
Reason
Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.
Root cause
Process control
Action
Zimmer sent an " URGENT MEDICAL DEVICE RECALL - LOT SPECIFIC " notifications dated September 2014, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
Quantity
618 units
Distribution
Worldwide Distribution - USA (nationwide) including the states of NY, OR, NC, OH, TX, FL, CO, IN, TN, PA, CA, KS, MO, WI, VA, MN, NJ, UT, MA, GA, MI, OK, AZ, MS, IL and AL., and the countries of Korea, Singapore, Canada and Japan
Lot, serial or model codes
Part Number 42-5299-001-00 Lots: 56572679; 56573170; 56572918; 56573378; 56572975; 56573379; 56572978; 56574223; 56572981; 56574291; 56572982; 56574292; 56573071; 56574316; 56573115; 56574317; 56573116; 56574318; 56573154.
510(k) numbers
["K113369"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register