Skip to the content

FDA Medical Device Recalls (openFDA)

OLEA MEDICAL: Functional MR V1.0

Recall number
Z-0122-2024
Recalling firm
OLEA MEDICAL
Product
Functional MR V1.0
Status
Open, Classified
Initiated
2023-08-10
Posted
2023-10-18
Terminated
not on file
Reason
When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.
Root cause
Software design
Action
On 8/10/23, field safety notices were emailed to customers who were told to take the following actions: Advice for the user to detect if the bug occurred: This defect can be detected within the Functional MR V1.0 user interface by displaying the exported volume on an anatomical sequence. To detect the issue, the VOI shall be dragged and dropped from the data browser on an anatomical sequence, which will display a notification, indicating the original VOI s name. Outside of Functional MR V1.0 user interface, the user may check the DICOM tag "Segment Label (0062,0003)" in the viewer used, as this DICOM tag contains the correct VOI s name. Workaround: This problem can be overcome when exporting VOIs individually and has no impact on exporting VOIs merged with an anatomical map. New software will be available in October 2023. Complete and return the reply form via email to qa-ra@olea-medical.com
Quantity
2
Distribution
US: MI, MA
Lot, serial or model codes
UDI-DI: EOLEFMRIV1P00, Software Versions: V1.0, V1.0 SP1, V1.0 SP2, V1.0 SP3
510(k) numbers
["K221426"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3007675959
Firm city
LA CIOTAT
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register