FDA Medical Device Recalls (openFDA)
Centurion Medical Products Corporation: Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR
- Recall number
- Z-0121-2020
- Recalling firm
- Centurion Medical Products Corporation
- Product
- Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR
- Status
- Terminated
- Initiated
- 2019-08-28
- Posted
- not on file
- Terminated
- 2020-07-23
- Reason
- A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
- Root cause
- Process control
- Action
- Centurion Medical Products issued Notification via Certified Mail, Return Receipt on 8/28/19 to customers who received affected product. Letter states reason for recall, helath risk and action to take: identify inventory within your possession and cease use immediately. Complete the enclosed accountability record, whether or not you have product, and fax or email at your earliest convenience to the attention of Lisa Carpenter as directed. Upon receipt of your response Centurion Representative will contact you regarding retrieval of the product from your facility, if required. Forward this recall information to any facilities to whom this product may have been further distributed, instructing them to cease use immediately. Questions: Kara A. Kowalczy Director of Quality Assurance(517) 546-5400 Ext.1122. Centurion mailed four (4) additional customer letters on 9/27 /2019 via Certified Mail, Return Receipt.
- Quantity
- 204 each
- Distribution
- Nationwide
- Lot, serial or model codes
- Batch/Lot: 2019061201
- 510(k) numbers
- ["K890897"]
- PMA numbers
- not on file
- Product code
- HFX
- Firm FEI
- 3008403546
- Firm city
- Williamston
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28