Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc: Zimmer brand Knowles pin with 5/32 in. threaded diameter, 1/8 in. shaft diameter and 3 1/2 inches in length. Catalog # 281-01-05.

Recall number
Z-0120-04
Recalling firm
Zimmer Inc
Product
Zimmer brand Knowles pin with 5/32 in. threaded diameter, 1/8 in. shaft diameter and 3 1/2 inches in length. Catalog # 281-01-05.
Status
Terminated
Initiated
2003-10-15
Posted
2004-07-20
Terminated
2004-05-20
Reason
The pin may fail because the metal used to make the pin does not meet the firm's hardness specification.
Root cause
Other
Action
All affected U.S. accounts were sent a letter dated October 15, 2003 by certified mail, return receipt requested. The seven foreign initial consignee accounts were notified via electronic mail. Customers were asked to return non-implanted product and to advise the implanting surgeons of the problem so that they may take appropriate action in regard to their patients.
Quantity
not on file
Distribution
The majority of the pins were distributed within the United States. Distribution also occured in seven foreign countries; Taiwan, Canada, Japan, Korea, Belgium, Australia and Finland.
Lot, serial or model codes
Lots 00330600, 54162100, 54382300, 55039300, 55147500, 55236100, 55417000, 60013747 and 60019270.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JDW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register