Skip to the content

FDA Medical Device Recalls (openFDA)

B BRAUN MEDICAL INC: PERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFCS. Product Description: CE17TKFCS CONT EPID TRAY W/17G TUOHY-LF.

Recall number
Z-0119-2026
Recalling firm
B BRAUN MEDICAL INC
Product
PERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Catheter. Product Code: CE17TKFCS. Product Description: CE17TKFCS CONT EPID TRAY W/17G TUOHY-LF.
Status
Open, Classified
Initiated
2025-09-18
Posted
2025-10-09
Terminated
not on file
Reason
Potential for the lid of the catheter connector to be in the incorrect position.
Root cause
Process control
Action
B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.
Quantity
21,270 units
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Lot, serial or model codes
REF: 332097; Product Code: CE17TKFCS; UDI-DI (Primary): 04046955899470; UDI-DI (Unit of Use): 04046955899487; Lot Numbers: 0062013051, 0062010532, 0062010533, 0062010534, 0062010535, 0062010536, 0062010555, 0062013050, 0062013053, 0062013052. Expiration Date: 07/31/2026.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OGE
Firm FEI
3013764800
Firm city
Allentown
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register