Zimmer Inc.: NexGen Complete Knee Solution Stemmed Tibial Component Precoat Size 6, REF 00-5980-047-02, Sterile, Zimmer, Warsaw, IN. Intended for mating with a UHMWPE articular surface and Zimaloy Cobalt-Chromium-Molybdenum alloy femoral component and is indicated for cemented use only.
- Recall number
- Z-0119-2011
- Recalling firm
- Zimmer Inc.
- Product
- NexGen Complete Knee Solution Stemmed Tibial Component Precoat Size 6, REF 00-5980-047-02, Sterile, Zimmer, Warsaw, IN. Intended for mating with a UHMWPE articular surface and Zimaloy Cobalt-Chromium-Molybdenum alloy femoral component and is indicated for cemented use only.
- Status
- Terminated
- Initiated
- 2010-01-26
- Posted
- 2010-10-21
- Terminated
- 2010-10-23
- Reason
- Zimmer had determined that two lots of these tibial implants have incorrectly positioned or missing flange plugs that were not seated in the device upon receipt to the customer. A missing flange plug could lead to a delay of surgery or possibility of wear leading to more surgery.
- Root cause
- Employee error
- Action
- The firm sent an URGENT MARKET WITHDRAWAL to its sales force and distributors on 1/26/2010. The letter identifies the affected product and explains the reason for the Market Withdrawal. The letter also explains the clinical implications and instructs consignees to locate any unused devices and quarantine them immediately. A physical count should be carried out of all affected devices and recorded on the Inventory Return Certification Form that's included with the letter. A copy of the letter should be faxed to Zimmer, Inc. and the affected product should be returned along with the form to the firm.
- Quantity
- 61 untis
- Distribution
- Worldwide Distribution -- US, Canada, Singapore Australia, and India.
- Lot, serial or model codes
- Lot Number: 61246477.
- 510(k) numbers
- ["K933785"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28