Philips Healthcare Informatics, Inc.: iSite PACS (Picture Archiving and Communication System), software versions 3.6.28.x and 4.1.x. The product is an image management system (software package) used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Device is intended for use by trained professionals.

Recall number
Z-0118-2009
Recalling firm
Philips Healthcare Informatics, Inc.
Product
iSite PACS (Picture Archiving and Communication System), software versions 3.6.28.x and 4.1.x. The product is an image management system (software package) used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. Device is intended for use by trained professionals.
Status
Terminated
Initiated
2008-08-29
Posted
2008-10-28
Terminated
2010-11-30
Reason
Two defects have been identified: 1) potential to display a patient on the canvas page that is different from the patient whose images are displayed on diagnostic monitors when using Conference Presentation States and when opening two or more studies from a folder or an exam worklist and 2) potential to miscalculate measurements when Pixel Spacing and Imager Pixel Spacing DICOM tag values are bot
Root cause
Software design
Action
A notification letter (Urgent Product Correction) dated 8/29/08 was sent via certified mail to all customers alerting them of the issue, requiring posting of notification. Philips Healthcare Informatics, Inc. strongly recommend all customers advise their iSite Radiology users not to open multiple patient exams simultaneously from a folder or worklist using Conference Presentation States. A software correction is being worked on, and should be available by end of Q3 2008 for version 4.1.45.2. A mandatory upgrade is required for all customers who are currently using iSite version 4.1.45.2 and former versions. For specific questions regarding these issues, contact the Customer Advocate (CARE) or Customer Care at Philips Healthcare Informatics, Inc. at 1-877-328-2808. In your communication with Customer Care at Philips Healthcare Informatics, Inc. , please refer to this notification and associated FCO 22945704-08292008-004.
Quantity
27 units.
Distribution
Worldwide Distribution including USA (MN, CO and PA) and countries of Italy, Germany, Netherlands, Sweden, France, Portugal, Switzerland, Thailand, Japan and Australia.
Lot, serial or model codes
Software versions 3.6.28.x and 4.1.x.
510(k) numbers
["K042292","K063267"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3002971142
Firm city
Foster City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register