Skip to the content

FDA Medical Device Recalls (openFDA)

Spineology, Inc.: The Graft Tube is an optional instrument which, as designed, can be filled with flowable graft material, then advanced through the Access Portal. A Push Rod is then used to facilitate the manual deployment of the graft material out of the Graft Tube and adjacent to where the implant is to be placed, as supported within the Surgical Technique Guide.

Recall number
Z-0117-2025
Recalling firm
Spineology, Inc.
Product
The Graft Tube is an optional instrument which, as designed, can be filled with flowable graft material, then advanced through the Access Portal. A Push Rod is then used to facilitate the manual deployment of the graft material out of the Graft Tube and adjacent to where the implant is to be placed, as supported within the Surgical Technique Guide.
Status
Open, Classified
Initiated
2024-09-16
Posted
2024-10-17
Terminated
not on file
Reason
Possible detachment of the Graft Tube shaft from the Graft Tube funnel head when it is not used as instructed by our surgical technique.
Root cause
Under Investigation by firm
Action
Spineology began notification of medical institutions on 09/16/2024, direct sales representatives on 09/18/2024 and distributors on 09/19/2024 via letter. Instructions included to notify appropriate staff of the action and to complete and return the response form. Spineology will be performing a correction to update the surgical technique labeling and a removal will be conducted to replace current Graft Tubes within field inventory. Distributors were instructed to notify customers if further distributed.
Quantity
190 units
Distribution
Pending
Lot, serial or model codes
UDI-DI: M7403120056, Catalog Number: 312-0056, part Number: 33-09-14. Lot Numbers: CH20003, BE22003
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003179719
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source