Beckman Coulter Inc.: Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116

Recall number
Z-0117-2022
Recalling firm
Beckman Coulter Inc.
Product
Power Processor 3K, 3K High Speed Stockyards Module Nameplate: PP 3K Stockyard (Refrig) 9707 REF A88116
Status
Open, Classified
Initiated
2021-07-28
Posted
2021-10-14
Terminated
not on file
Reason
Missing splash guard which may expose users to hazardous materials. .
Root cause
Process control
Action
During the week of 08/30/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via its Field Action Notification process. On 09/16/2021, the Recalling Firm communicated a updated/revised "URGENT MEDICAL DEVICE RECALL" Notification letter which informed customer that, as a result of internal investigations, the Recalling Firm has discovered that the track-level splash guard has not been consistently provided to all customers. This guard is required for safe operation of the Power Processor system and could potentially cause user injuries if the splash guard is not in place. Customer are being informed that a Recalling Firm Field Service representative will check their system for the splash guard. if the splash guard is not present, the Field Service representative will order and install it. Customers are instructed to: 1. Continue to follow all safety warnings and system operation information provided in the Instructions for Use. 2. Share this information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation. 3. If customers have forwarded any of the affected products to another laboratory, provide a copy of this notification to them. 4. To confirm receipt of this Notification, customer are asked to respond within 10 days in one of the following ways: -Electronically, if you received this communication via email. -Manually, complete and return the enclosed Response Form. Any questions, contact the Recalling Firm's Customer Support Center or their local representative via website http://www.beckmancoulter.com Update 12/15/2021: On 12/13/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter informing of additional products being added to the original Recall initiated in 08/2021 for the missing splash guards. A Recalling Firm's Representative will check the systems and if a splash guard is not present, they will order and install it. In addition, the Recalling Firm is a
Quantity
Total Units=271 units [258 units (3K and 3K HS); 13 units (5K added 12/15/2021)]
Distribution
U.S. Nationwide Distribution in the states of: CA, GA, IA, ID, IL, IN, LA, MA, MI, MS, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WA, and WV O.U.S.: Canada, China, Spain, France, United Kingdom, Israel, Italy, South Korea, Kuwait, Lebanon, Mexico, Malaysia, Netherlands, Portugal, Qatar, Saudi Arabia, Thailand, Turkey, and Taiwan. Additional O.U.S. added 12/15/2021 - South Africa
Lot, serial or model codes
Catalog Number: A88116 - 3K HS Refrigerated Stockyard UDI: 15099590591960 Serial Numbers: 90HF1304006; 90HF1306009; 90HF1303005; 90HF1508005; 90HF1302005; 90HF1610005; 90HF1806001; 90HF1209005; 90HF1303007; 90HF1708006; 90HF1205001; 90HF1805002; 90HF1402005; 90HF1503001; 90HF1104002; 90HF1208004; 90HF1208005; 90HF1203001; 90HF1012004; 90HF1101002; 90HF1702003; 90HF1402004; 90HF1303008; 90HF1210001; 90HF1207001; 90HF1305007; 90HF1305004; 90HF1610001; 90HF1505002; 90HF1307005; 90HF1610002; 90HF1508003; 90HF1210002; 90HF1306001; 90HF1206002; 90HF1208002; 90HF1305008; 90HF1009002; 90HF1101001; Catalog Number: 6915556 - Power Processor 3K Stockyard (Stockyard 3000/R Kit for PWR PROC) Serial Numbers: 90AM0805001; 90AM0902009; 90AM0902005; 90AM0806003; 90AM0801003; 90AM0802002; 90AM0803001; 90AM0912002; 90AM1003002; 90AM0905010; 90AM0801004; 90AM06007008; 90AM060703; 90AM060704; 90AM0710002; 90AM0905007; 90AM0710008; 90AM0903002; 90AM0808002; 90AM0711004; 90AM0711005; 90AM0810001; 90AM0901002; 90AM0905006; 90AM0904004; 90AM0904005; 90AM0803002; 90AM0710004; Amended - additional Products and Codes (12/15/2021): B05520 - Stockyard, 5K HS, Refrigerated Serial Numbers: 90HG1104002; 90HG1101001; 90HG1105002; 90HG1104001; 90HG1101003; 90HG1101002; 90HG1202004; 90HG1102005; 90HG1101004; 90HG1206002; A88115 - 5K HS Refrig Stockyard Serial Numbers: 90HG0911001 90HG0911003 90HG0911002
510(k) numbers
["K110413"]
PMA numbers
not on file
Product code
CDD
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register