FDA Medical Device Recalls (openFDA)
Straumann USA, LLC: Straumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery.
- Recall number
- Z-0117-2015
- Recalling firm
- Straumann USA, LLC
- Product
- Straumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery.
- Status
- Terminated
- Initiated
- 2014-09-19
- Posted
- 2014-10-23
- Terminated
- 2015-03-13
- Reason
- Assembled with Narrow Connect (NC) transfer piece instead of Regular Connect (RC)
- Root cause
- Packaging process control
- Action
- Strauman notified consignees of this Field Corrective Action by letter dated September 19, 2014 and delivered via hand by Straumann Representative. Accounts are requested to return inventory. Contact (978) 747-2514 for assistance.
- Quantity
- 6 units
- Distribution
- US Distribution in states of: FL and MA.
- Lot, serial or model codes
- Lot number: GY030 Exp. 2019-04
- 510(k) numbers
- ["K140878"]
- PMA numbers
- not on file
- Product code
- DZE
- Firm FEI
- 1000121052
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28