Skip to the content

FDA Medical Device Recalls (openFDA)

Straumann USA, LLC: Straumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery.

Recall number
Z-0117-2015
Recalling firm
Straumann USA, LLC
Product
Straumann Bone Level Tapered Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021 .5308 Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery.
Status
Terminated
Initiated
2014-09-19
Posted
2014-10-23
Terminated
2015-03-13
Reason
Assembled with Narrow Connect (NC) transfer piece instead of Regular Connect (RC)
Root cause
Packaging process control
Action
Strauman notified consignees of this Field Corrective Action by letter dated September 19, 2014 and delivered via hand by Straumann Representative. Accounts are requested to return inventory. Contact (978) 747-2514 for assistance.
Quantity
6 units
Distribution
US Distribution in states of: FL and MA.
Lot, serial or model codes
Lot number: GY030 Exp. 2019-04
510(k) numbers
["K140878"]
PMA numbers
not on file
Product code
DZE
Firm FEI
1000121052
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register