Advanced Bionics Corporation: HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument.

Recall number
Z-0117-2013
Recalling firm
Advanced Bionics Corporation
Product
HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument.
Status
Terminated
Initiated
2007-09-14
Posted
2012-10-22
Terminated
2012-10-22
Reason
Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.
Root cause
Nonconforming Material/Component
Action
The firm, Advanced Bionics, sent a " RECALL NOTIFICATION" letter dated September 14, 2007 to all their customers who purchased the HiRes 90K devices with Helix Electrodes Model CI 1400-02H. The letter identified the product, problem, and actions to be taken. The customers were instructed to immediately locate and return the affected product to Advanced Bionics. The firm's Service representative will contact the customers within 48 hours to retrieve the affected product and send a replacement as needed. The customers were also instructed to complete and return the enclosed Notification Acknowledgement form via fax to: (661) 362-1511. If you have any questions regarding this letter, call Advanced Bionics at (877) 829-0026 between the hours of 5 AM and 5 PM Pacific Time, Monday through Friday.
Quantity
17 units
Distribution
Worldwide distribution: USA (nationwide) including state: IL and country of: Europe.
Lot, serial or model codes
Serial Numbers:  351323, 251153, 350989, 351226, 350196, 351136, 350140, 350201, 351261, 351264, 351266, 351269, 351412, 351415, 351429, 350157, 351303.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MCM
Firm FEI
1000220815
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register