Philips Healthcare Inc.: Philips BV Endura Fluoroscopic Image Intensified X-ray Catalog Number: 718074 Product Usage: Fluoroscopic x-ray system, image-intensified Fluoroscopic x-ray system.

Recall number
Z-0117-2012
Recalling firm
Philips Healthcare Inc.
Product
Philips BV Endura Fluoroscopic Image Intensified X-ray Catalog Number: 718074 Product Usage: Fluoroscopic x-ray system, image-intensified Fluoroscopic x-ray system.
Status
Terminated
Initiated
2011-10-04
Posted
2011-10-27
Terminated
2017-01-12
Reason
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Root cause
Nonconforming Material/Component
Action
Philips Healthcare Inc., sent an "URGENT - Field Safety Notice" dated September 30, 2011 to all affected customers. The notice identifies the following: what the problem is and under what circumstances it can occur; the actions that should be taken by the customer/user in order to prevent risks for patients or users and the actions planned by Philips Medical Systems to correct the problem. The notice contains important information for the continued safe and proper use of the equipment and customers were advised to retain a copy with the equipment Instruction for Use. The notice stated that Philips will replace the front plates for systems on the Unit Affected List by the end of June 2012. The notice instructed customers to contact Philips Healthcare representative for further information or support.
Quantity
12 units in the US
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Australia, Austria, Belgium, Bharain, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Netherlands, Norway, Portugal, Russian Fed., Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
Lot, serial or model codes
Shipped from June~2010 to 26 Jul 2011 S/N: 1413 1438 1492 1504 1567 1567 1620 1633 1634 1691 1765 1800
510(k) numbers
["K010435"]
PMA numbers
not on file
Product code
JAA
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register