Philips Healthcare Inc.: Philips BV Pulsera Fluoroscopic Image Intensified X-ray Catalog Number: 718095 Product Usage: Fluoroscopic x-ray system, image-intensified Fluoroscopic x-ray system.

Recall number
Z-0116-2012
Recalling firm
Philips Healthcare Inc.
Product
Philips BV Pulsera Fluoroscopic Image Intensified X-ray Catalog Number: 718095 Product Usage: Fluoroscopic x-ray system, image-intensified Fluoroscopic x-ray system.
Status
Terminated
Initiated
2011-10-04
Posted
2011-10-27
Terminated
2017-01-12
Reason
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Root cause
Nonconforming Material/Component
Action
Philips Healthcare Inc., sent an "URGENT - Field Safety Notice" dated September 30, 2011 to all affected customers. The notice identifies the following: what the problem is and under what circumstances it can occur; the actions that should be taken by the customer/user in order to prevent risks for patients or users and the actions planned by Philips Medical Systems to correct the problem. The notice contains important information for the continued safe and proper use of the equipment and customers were advised to retain a copy with the equipment Instruction for Use. The notice stated that Philips will replace the front plates for systems on the Unit Affected List by the end of June 2012. The notice instructed customers to contact Philips Healthcare representative for further information or support.
Quantity
134 units in the US
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Australia, Austria, Belgium, Bharain, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Netherlands, Norway, Portugal, Russian Fed., Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
Lot, serial or model codes
Shipped from May 2010 to 26 Jul 2011. S/N: 774 1463 1492 1568 1572 1597 1598 1599 1600 1601 1603 1604 1605 1606 1607 1608 1609 1615 1616 1620 1634 1648 1649 1650 1654 1655 1656 1657 1658 1659 1660 1661 1662 1663 1677 1679 1681 1682 1703 1709 1710 1711 1712 1713 1726 1727 1731 1732 1734 1735 1736 1741 1742 1743 1744 1745 1746 1758 1783 1784 1797 1818 1819 1826 1827 1831 1848 1857 1858 1859 1865 1866 1867 1868 1869 1872 1885 1894 1898 1900 1903 1908 1919 1922 1930 1931 1936 1937 1938 1946 1950 1951 1955 1956 1959 1965 1966 1967 1968 1970 1979 1980 1981 1982 1983 1984 1986 1987 1988 1989 1990 1991 1992 1998 1999 2000 2003 2005 2012 2016 2017 2018 2020 2021 2022 2025 2042 2043 2044 2046 2064 2074 2075
510(k) numbers
["K010435"]
PMA numbers
not on file
Product code
JAA
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register