FDA Medical Device Recalls (openFDA)
S.A.E. AFIKIM: Breeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It is intended to provide increased mobility for elderly and/or disabled person and it is also an aid to independent living. Products can be identified by their color; the recalled products are either metallic blue or metallic silver. The products have two rear wheels and have either one or two wheels in the front. The scooters have an adjustable swivel seat and a front storage basket; some scooters may be fitted with an optional larger rear basket.
- Recall number
- Z-0115-2014
- Recalling firm
- S.A.E. AFIKIM
- Product
- Breeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It is intended to provide increased mobility for elderly and/or disabled person and it is also an aid to independent living. Products can be identified by their color; the recalled products are either metallic blue or metallic silver. The products have two rear wheels and have either one or two wheels in the front. The scooters have an adjustable swivel seat and a front storage basket; some scooters may be fitted with an optional larger rear basket.
- Status
- Terminated
- Initiated
- 2013-09-27
- Posted
- 2013-10-30
- Terminated
- 2014-07-23
- Reason
- Breeze C Scooter, motorized 3-wheeled electric scooter for elderly people, is recalled because it may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.
- Root cause
- Process control
- Action
- Afikim Electric Vehicles sent an URGENT: MEDICAL DEVICE RECALL letter, dated September 27, 2013, to all affected customers and distributors. . Customers are advised to remove the ignition key and contact Afikim Electric Vehicles via telephone Monday through Friday at 1-800-809-3010 between the hours of 9AM and 5PM (EST). Consumers may also contact the company via e-mail at dov@afiscooters.com. Customers are instructed to return the Recall Response Form by fax it to 800-466-1171 or email to ofer@afiscooters.com. Consumers should report any adverse event or quality problems to FDA's MedWatch Adverse Event Reporting program either by online, fax, or regular mail.
- Quantity
- 69 units
- Distribution
- Nationwide Distribution: US including the states of California, Florida, Georgia, Illinois, Maine, Maryland, Massachusetts, Michigan, Minnesota, New Jersey, New York, Ohio, Oklahoma, Pennsylvania, Texas, Virginia, and Wisconsin.
- Lot, serial or model codes
- Serial Numbers: 10390024, 10390025, 10390029, 10390030, 10390034, 10390037, 10490100, 10490108, 10490111, 10490113, 10490119, 10490121, 10490124, 11390058, 11390060, 11390061, 11390069, 11390074, 11390095, 11390101, 11490111, 11490147, 11490176, 11490189, 11490277, 11490304, 11490305, 11490425, 11490426, 11490433, 11490435, 11490445, 12390096, 12390097, 12390098, 12390099, 12390101, 12390102, 12390103, 12390104, 12390105, 12390106, 12390107, 12390117, 12390120, 12490169, 12490170, 12490172, 12490173, 12490174, 12490175, 12490176, 12490177, 12490178, 12490179, 12490180, 12490181, 12490182, 12490183, 12490184, 12490185, 12490186, 12490187, 12490188, 12490189, 12490190, 12490191, 12490192, and 12490193.
- 510(k) numbers
- ["K102304"]
- PMA numbers
- not on file
- Product code
- INI
- Firm FEI
- 1000646143
- Firm city
- Afikim
- Firm state
- not on file
- Firm country
- Israel
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28