FDA Medical Device Recalls (openFDA)
Radiometer America Inc: Radiance Data Management System, Software Version 2.60, modification for the ABL800 Flex Blood Gas Analyzer
- Recall number
- Z-0115-2007
- Recalling firm
- Radiometer America Inc
- Product
- Radiance Data Management System, Software Version 2.60, modification for the ABL800 Flex Blood Gas Analyzer
- Status
- Terminated
- Initiated
- 2006-07-06
- Posted
- 2006-10-28
- Terminated
- 2011-11-18
- Reason
- Incorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.
- Root cause
- Other
- Action
- The recalling firm initially contacted their customers and informed them of the software problem and the upcoming field correction for the software, via telephone on 7/6/2006. A follow-up letter dated 7/21/2006, was also sent to the customers.
- Quantity
- 5 units
- Distribution
- Worldwide-The six USA units (equpped with Radiance software) were distributed to a total of two (2) U.S. customers: University of Missouri Hospitals and Clinics in Columbia, MO (4 units) and Vail Valley Medical Center, Vail, CO (2 units). World wide accounts pending submission. Three (3) Canadian customers. They are: 1) Seven Oaks General Hospital MB Canada; and 2) St. Mary''s General Hospital, 911 Queens Blvd, Ontario Canada; and 3) University of Alberta Hospital, 84440 112th Street, Alberta, Canada.
- Lot, serial or model codes
- Catalog Numbers: 914-418 and 914-426
- 510(k) numbers
- ["K050869"]
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 1523456
- Firm city
- Westlake
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28