Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Craniomaxillofacial Division: Disposable Paper Filter for Sterilization Containers Paper, Medical Accessory, 100 DISP, Paper F, Steri.Cases, New Shape, Manufactured and distributed by Stryker Leibinger GmbH & Co KG, Distributed in the USA by: Stryker Orthopedics, NON STERILE.

Recall number
Z-0114-2016
Recalling firm
Stryker Craniomaxillofacial Division
Product
Disposable Paper Filter for Sterilization Containers Paper, Medical Accessory, 100 DISP, Paper F, Steri.Cases, New Shape, Manufactured and distributed by Stryker Leibinger GmbH & Co KG, Distributed in the USA by: Stryker Orthopedics, NON STERILE.
Status
Terminated
Initiated
2015-09-17
Posted
2015-10-08
Terminated
2016-02-26
Reason
Stryker has become aware that the inhomogeneity of the filter paper might potentially compromise the ability of the filter to maintain a sterile barrier during post-sterilization shelf-life.
Root cause
Device Design
Action
An Urgent Medical Device Recall letter has been sent to the customers. The affected product was identified, as well as the reason for the recall and risk to health. The letter states that customers should inform users of the Medical Device Removal and pass the notice to all appropriate individuals within the organization; scrap all affected products available at the location or return it to Stryker Craniomaxillofacial, Attention: Recall Coordinator, at the address provided. If there are questions regarding the impacted product, don't hesitate to Contact Stryker Customer service and refer to PFA #2015-032. Complete and sign the enclosed Business Reply Form and fax or copy to: (877) 648-7114 or e-mail a copy to CMF-custserv@stryker.com.
Quantity
915 packs
Distribution
Worldwide distribution -- US, Canada, France, Germany, Italy, Japan, Norway, Sweden, Switzerland, and the Netherlands.
Lot, serial or model codes
Part Number:  29-10911 Disposable Paper Filter for Sterilization Container, 30x29cm, new type (P/100);  29-10912 Disposable Paper Filter for Sterilization Container, 60x 29cm, old type (P/100), obsolete (Part not distributed in the US);  29-10913 Dispoable Paper Filter for Sterilization Container, 60x29cm new type (P/100) (Part not distributed in the US);  29-10915 Disposable Paper Filter, 4-holes, for Coiner (Part not distributed in the US)
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSM
Firm FEI
3005101424
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register