Ge Healthcare: GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.

Recall number
Z-0113-2008
Recalling firm
Ge Healthcare
Product
GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.
Status
Terminated
Initiated
2007-07-06
Posted
2007-11-07
Terminated
2008-10-21
Reason
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Root cause
Software design
Action
GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
Quantity
521 Units
Distribution
Nationwide.
Lot, serial or model codes
Serial Numbers: D00210, D00223, D00225, D00239, D00241, D00293, D00326, D00347, D00415, D00425, D00435, D00445, D00455, D00456, D00459, D00460, D00465, D00467, D00468, D00469, D00480, D00481, D00484, D00485, D00487, D00490, D00491, D00493, D00495, D00502, D00503, D00505, D00512, D00516, D00517, D00523, D00525, D00531, D00532, D00537, D00539, D00541, D00542, D00543, D00546, D00548, D00550, D00551, D00552, D00554, D00558, D00560, D00561, D00562, D00563, D00564, D00565, D00566, D00567, D00568, D00569, D00570, D00571, D00572, D00573, D00574, D00575, D00576, D00577, D00579, D00580, D00581, D00582, D00625, D00627, D00628, D00632, D00633, D00636, D00638, D00639, D00640, D00649, D00652, D00653, D00655, D00665, D00666, D00668, D00108, D00123, D00134, D00139, D00168, D00173, D00176, D00178, D00182, D00185, D00186, D00187, D00189, D00192, D00193, D00194, D00195, D00198, D00199, D00200, D00203, D00209, D00214, D00220, D00224, D00226, D00227, D00228, D00230, D00231, D00232, D00233, D00236, D00242, D00243, D00244, D00246, D00249, D00252, D00254, D00258, D00282, D00291, D00292, D00295, D00296, D00297, D00299, D00300, D00302, D00303, D00304, D00305, D00306, D00307, D00310, D00312, D00314, D00316, D00317, D00318, D00319, D00327, D00330, D00335, D00336, D00337, D00339, D00340, D00343, D00344, D00346, D00349, D00350, D00351, D00352, D00354, D00355, D00356, D00359, D00361, D00362, D00364, D00366, D00367, D00372, D00374, D00375, D00388, D00391, D00392, D00393, D00395, D00396, D00397, D00398, D00399, D00402, D00403, D00404, D00405, D00406, D00411, D00413, D00418, D00420, D00421, D00422, D00423, D00426, D00451, D00180, D00379, D00387, D00430, D00433, D00444, D00454, D00464, D00470, D00476, D00497, D00509, D00513, D00522, D00526, D00540, D00544, D00545, D00547, D00549, D00553, D00555, D00556, D00557, D00559, D00585, D00586, D00103, D00155, D00166, D00250, D00251, D00255, D00325, D00331, D00595, D00584, D00280, D00342, D00369, D00148, D00213, D00301, D00428, D00601, D00602, D00603, D00604, D00324, D00151, D00471, D00519, D00593, D00466, D00494, D00659, D00432, D00473, D00321, D00591, D00599, D00153, D00217, D00222, D00248, D00437, D00514, D00322, D00449, D00474, D00475, D00478, D00504, D00507, D00508, D00521, D00528, D00529, D00533, D00534, D00583, D00634, D00635, D00641, D00644, D00645, D00663, D00126, D00128, D00240, D00272, D00277, D00279, D00332, D00341, D00407, D00419, D00486, D00424, D00499, D00288, D00308, D00394, D00483, D00143, D00117, D00138, D00150, D00165, D00167, D00183, D00218, D00245, D00253, D00259, D00261, D00264, D00315, D00328, D00378, D00383, D00385, D00431, D00462, D00613, D00618, D00121, D00141, D00154, D00170, D00171, D00190, D00234, D00256, D00267, D00311, D00313, D00461, D00145, D00146, D00196, D00201, D00205, D00206, D00207, D00257, D00263, D00266, D00268, D00410, D00438, D00439, D00458, D00479, D00488, D00489, D00496, D00511, D00518, D00520, D00527, D00538, D00594, D00596, D00598, D00600, D00605, D00621, D00660, D00661, D00179, D00202, D00204, D00208, D00215, D00219, D00221, D00237, D00260, D00265, D00271, D00275, D00276, D00281, D00285, D00286, D00287, D00353, D00358, D00360, D00365, D00429, D00309, D00334, D00368, D00477, D00482, D00149, D00152, D00289, D00290, D00323, D00401, D00211, D00144, D00181, D00191, D00269, D00472, D00588, D00589, D00161, D00162, D00164, D00229, D00329, D00370, D00371, D00390, D00427, D00160, D00212, D00283, D00320, D00122, D00137, D00156, D00377, D00409, D00434, D00446, D00447, D00450, D00452, D00453, D00463, D00492, D00506, D00510, D00515, D00524, D00530, D00535, D00578, D00587, D00590, D00592, D00597, D00606, D00607, D00608, D00609, D00610, D00611, D00612, D00614, D00615, D00616, D00617, D00619, D00620, D00622, D00623, D00624, D00626, D00629, D00630, D00631, D00642, D00643, D00647, D00650, D00651, D00654, D00656, D00658, D00662, D00664, D00667, D00669, D00670, D00672, D00101, D00105, D00107, D00113, D00125, D00127, D00131, D00132, D00135, D00136, D00158, D00163, D00175, D00188, D00278, D00294.D00333, D00357, D00373, D00376, D00381, D00382, D00384, D00386, D00389, D00400, D00408, D00412, D00414, D00416, D00417, D00436, D00440, D00441, D00442, D00443, D00498, D00500, D00501.  Version 6.2.0 was distributed 3/24/07 and version 6.2.1: 4/10/07
510(k) numbers
["K061682"]
PMA numbers
not on file
Product code
IYO
Firm FEI
3005860720
Firm city
Wauwatosa
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register