Stryker Instruments Div. of Stryker Corporation: Sagittal Blade 18.0X0.97X90MM, Model Number: 6118-097-090 Product Usage: Heavy duty Sagittal Blades are used to cut bone and bone related tissue in a variety of orthopedic procedures. Typical procedures include, total-knee, uni-compartmental knee, hip, shoulder and ankle procedures. These blades can be used with the following handpieces; 4208-000-000 (SYSTEM 5 SAGITTAL SAW) 6208-000-000 (SYSTEM 6 SAG SAW) 6298-000-000 (ES6 SAGITTAL SAW) 7208-000-000 (SYSTEM 7 SAG SAW)

Recall number
Z-0112-2017
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Sagittal Blade 18.0X0.97X90MM, Model Number: 6118-097-090 Product Usage: Heavy duty Sagittal Blades are used to cut bone and bone related tissue in a variety of orthopedic procedures. Typical procedures include, total-knee, uni-compartmental knee, hip, shoulder and ankle procedures. These blades can be used with the following handpieces; 4208-000-000 (SYSTEM 5 SAGITTAL SAW) 6208-000-000 (SYSTEM 6 SAG SAW) 6298-000-000 (ES6 SAGITTAL SAW) 7208-000-000 (SYSTEM 7 SAG SAW)
Status
Terminated
Initiated
2016-08-17
Posted
2016-10-18
Terminated
2017-03-29
Reason
Stryker Instruments is voluntarily recalling the Sagittal Blade 18.0X.97X90MM due to the potential for a sterile barrier breach in the product's packaging.
Root cause
Packaging
Action
Stryker Instruments is voluntarily recalling the Sagittal Blade 18.0X.97X90MM, product number 6118-097-090, lot number 16029017, due to a breach in the packaging causing a breach in sterility. There is a potential for a bone or soft tissue infection which may require medical or surgical intervention. Customer notices were issued via certified mail on 08/17/2016. Actions to be taken by the Customer/User: 1. Immediately review this Recall Notification. 2. Check all stock areas and/or operating room storage to determine how many Sagittal Blades from the affected lot (listed above) are at your facility. Quarantine and discontinue use of the recalled Sagittal Blades. 3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if customers dont have any affected product on hand. Note: Customer signature on the BRF indicates that you received and understand this Notification and have followed the instructions in the Notification. 4. If customers have further distributed this product, please forward the recall letters and the attached BRF to all affected locations. Please indicate each location on the BRF. 5. Fax the completed BRF to Stryker Instruments Regulatory Department at 866-521-2762, or scan and email a copy to kellyjo.davis@stryker.com. 6. If the BRF for customer facilities indicates that recalled product is currently on hand, a FedEx label will be emailed to you. This shipping label should be used to return recalled product. 7. Upon receipt of the recalled product, replacement product will be shipped to customer accounts. For questions regarding this recall, please contact Stryker Instruments: Kelly Jo Davis, 269-389-2921, kellyjo.davis@stryker.com
Quantity
225
Distribution
US Nationwide in the states of RI, WA, CA, CO, WI,FL, ID, ND, MD,TX Foreign: Australia, New Zealand, Netherlands, France, Italy
Lot, serial or model codes
Sagittal Blade 18.0X0.97X90MM, Model Number: 6118-097-090, Lot # 16029017
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWE
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register