DePuy Orthopaedics, Inc.: Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C. SIGMA knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included with the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Primary Knee Instruments, the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Revision Knee Instruments, and P.F.C. SIGMA High Performance (HP) Primary Knee Instruments

Recall number
Z-0111-2016
Recalling firm
DePuy Orthopaedics, Inc.
Product
Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C. SIGMA knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included with the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Primary Knee Instruments, the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Revision Knee Instruments, and P.F.C. SIGMA High Performance (HP) Primary Knee Instruments
Status
Terminated
Initiated
2015-09-08
Posted
2015-10-07
Terminated
2018-07-03
Reason
DePuy Orthopaedics, Inc. is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients.
Root cause
Device Design
Action
DePuy Synthes sent an Urgent Information Recall Notice to all affected customers. Instruments are not lot traceable in the system, so the company sent notices by e-mail on September 8, 2015, and US Distributors will hand deliver notices to all Medical Facilities that may have purchased the affected instrument. The US Distributor or US Sales Consultant will deliver the Medical Professional Notice and Reconciliation Form to the affected Medical Facilities. If assistance is needed, Sales Consultants will assist customers with inspection of all instruments, return of affected lots, receiving credit and returning the completed Medical Facility Reconciliation Forms to DePuy Orthopaedics, Inc. The affected instruments will be returned to the quarantine warehouse located at 700 Orthopaedic Drive, Warsaw, IN 46582. For questions regarding this recall call 574-371-4917. For questions regarding this recall call 574-267-8143.
Quantity
7488
Distribution
Worldwide Distribution - US including AL; AZ; CA; CO; CT; DC; FL; GA; HI; IA; IL; IN; KY; LA; MA; MD; ME; MI; MN; MO; MS; MT; NC; NE; NH; NJ; NV; NY; OH; OR; PA; RI; SC; TN; TX; UT; VA; VT; WA; WI; and WV and Internationally to ARGENTINA; AUSTRALIA; AUSTRIA; BELGIUM; BRAZIL; CANADA; CHILE; COLOMBIA; CZECH REPUBLIC; EGYPT; FRANCE; GERMANY; GREECE; INDIA; ISRAEL; ITALY; KOREA; Mexico; NEW ZEALAND; NORWAY; P R CHINA; PANAMA; PERU; POLAND; RUSSIA; SINGAPORE; SLOVENIA; SOUTH AFRICA; SWITZERLAND; UAE; UK; URUGUAY; VENEZUELA; and VIETNAM.
Lot, serial or model codes
SP2 IM Rod 400mm Instruments (96-6120) manufactured with 455 stainless steel (SS) or 17-4 SS; CATALOG NO. : 96-6120; GTIN NO.: 10603295246893; Lots: Manufactured with 455 Stainless Steel, Labeled Lot #s-C98CS4; DE5P34; DE5RP4; D95AN4; EB5FV4; FA4G94; DE1FA4; DF4H44; DG9LK4; EB5GH4; EC9JY4; EF4DJ4; FD8MP4; FH8JA4; DG9L64; DJ5E34; DK3E34; EJ7AP4; DK3FE4; EJ7A34; ES2G64; ES2HA4; FH8JX4; ES2HY4; EX5L44; C3JHN4; EX5MS4; C3JHS4; E2SD44; C3JHF4; C4GA54; C4GBT4; C4GCC4; C52F14; C52F74; C52GL4; C52GV4; C67N14; C98BF4; TBACC; TBACZ; FJ4E74; TBCOJ; & Manufactured with 17-4 Stainless Steel- Etched Lot #s- H0797; H0100; H0199; H0200; H0298; H0301; H0396; H0399; H0401; H0495; H0499; H0530; H0596; H0598; H0600; H0601; H0696; H0699; H0796; H0797; H0800; H0899; H0900; H0997; H0998; H1095; H1097; H1195; H1198; H1297; H1298; H1299
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register