Ethicon, Inc.: EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.

Recall number
Z-0111-2013
Recalling firm
Ethicon, Inc.
Product
EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.
Status
Terminated
Initiated
2012-07-02
Posted
2012-10-22
Terminated
2015-09-15
Reason
Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
Root cause
Employee error
Action
Ethicon sent a notification letter "Urgent Product Information EVICEL Fibrin Sealant (Human) All Lots and Product Codes" dated June 27, 2012, to all affected customers. Notifications were sent via UPS on July 2, 2012. The notification informed customers of the identified risks and included instructions for use. The notification included a business reply card consignees were requested to fill out and return. Customer were notified by a notification titled "Urgent Product Information Omrix Pressure Regulator All Product Codes" that was sent by Ethicon on August 1, 2012, as a follow up to the letter dated June 27, 2012. The notification informed consignees of the identified risks and included instructions for use. Customers were notified that a company representative would visit them to confirm receipt of the notification and to ensure the enclosed label has been affixed correctly on all Omrix Pressure Regulators. For further questions please call 1-877-384-4266.
Quantity
4800 pressure regulators distributed worldwide; 1500 distributed within the US
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Canada, European Union, Singapore, Israel, Mexico and Brazil.
Lot, serial or model codes
To include all lots and product codes for EVICEL and Omrix pressure regulators.
510(k) numbers
["K090162"]
PMA numbers
not on file
Product code
FMF
Firm FEI
2210968
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register