Ekos Corp: Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System consists of three main components, a single use Lysus Infusion Catheter System which is comprised of a Drug Delivery Catheter (DDC) and removable Ultrasound Core (USC), and a reusable EKOS PT-3 Control System. During use, fluids are delivered through the side-holes of the DDC. The USC is placed through the central lumen of the DDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS PT-3 Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Catheter Interface Cable with Pod connects the EKOS PT-3 Control Unit to the Lysus Infusion Catheter System.

Recall number
Z-0111-06
Recalling firm
Ekos Corp
Product
Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System consists of three main components, a single use Lysus Infusion Catheter System which is comprised of a Drug Delivery Catheter (DDC) and removable Ultrasound Core (USC), and a reusable EKOS PT-3 Control System. During use, fluids are delivered through the side-holes of the DDC. The USC is placed through the central lumen of the DDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS PT-3 Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Catheter Interface Cable with Pod connects the EKOS PT-3 Control Unit to the Lysus Infusion Catheter System.
Status
Terminated
Initiated
2004-11-22
Posted
2005-11-01
Terminated
2005-11-09
Reason
During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In one case, the USC fractured within the DDC when the user attempted to withdraw the USC after resistance was encountered
Root cause
Other
Action
On 11/22/04 the firm issued letters to those customers that had product. The letters advised of the problem and provided amended use instructions.The firm continued to ship product and on 3/10/05 the firm visited consignees, replacing the catheters.
Quantity
134 kits
Distribution
Kits were distributed to 6 hospitals in the US, 1 each located in AZ, OH, WA, FL, IL, and CA. The firm also distributed to a customer in Germany.
Lot, serial or model codes
Serial #s 040914006-004, 040914006-008, 040914007-006, 040914008-003, 040928003-002, 040928003-005, 040928003-009, 040928004-002, 040928005-004, 040928006-002, 040928006-003, 040928006-006, 041013005-004, 040702005-007, 040702005-008, 040719001-005, 040719001-006, 040913002-001, 040913002-002, 040913002-003, 040914007-002, 040914010-001, 040928004-005, 041013006-014, 040928003-012, 040928003-013, 041013005-006, 041013005-007, 041013006-011, 041013006-017, 040719002-003, 040322001-003, 040322001-006, 040322001-013, 040420001-001, 040420001-002, 040420001-004, 040420001-005, 040420001-006, 040420001-007, 040420001-008, 040420001-009, 040420001-010, 040420001-011, 040702005-001, 040702005-002, 040702006-001, 040702006-002, 040928004-004, 040928005-002, 040928005-005, 040928006-004, 040928006-005, 040702005-003, 040702005-004, 040702006-003, 040702006-004, 040719001-001, 040719001-002, 040719002-002, 040719003-001, 040719003-002, 040914007-005, 040914008-002, 040914009-003, 040928004-001, 040928006-009, 040928006-010, 040928007-002, 040702005-005, 040702005-006, 040702006-005, 040719001-003, 040719001-004, 040909003-001, 040909003-002, 040913002-004, 040913002-005, 040914006-007, 040914007-003, 040914007-004, 040914009-004, 040928003-001, 040928004-008, 040928005-006, 040928006-001, 040928007-001, 041013005-005, 041013006-012, 041013006-013, 041119002-001, 041119003-001, 040719001-007, 040914006-003, 040914007-001, 040914008-001, 040914009-001, 040928003-010, 040928004-006, 040928004-007, 041013004-001, 041013004-002, 041013004-003, 041013006-008, 041013006-015, 040420001-003, 040719001-008, 040719002-001, 040914006-001, 040914006-005, 040914006-006, 040914006-009, 040914006-010, 040914008-004, 040928003-003, 040928003-004, 040928003-011, 040928003-014, 040928004-003, 040928005-007, 040928005-008, 041013006-009, 041013006-016, 041013006-018, 050204002-001, 050204002-002, 050204002-003, 050204002-004, 050204002-005, 050209002-001, 050209002-002, 050209002-003, 050209002-004, 050209002-005   Catalog # & Part #: 500-52112, 3383-002; 500-52124, 3393-004; 500-52124, 3393-006; 500-52150, 3393-007
510(k) numbers
["K042456"]
PMA numbers
not on file
Product code
KRA
Firm FEI
3001627457
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register