Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD BACTEC Plus Aerobic/F Culture Vials, Catalog Number 442023

Recall number
Z-0110-2024
Recalling firm
Becton Dickinson & Co.
Product
BD BACTEC Plus Aerobic/F Culture Vials, Catalog Number 442023
Status
Open, Classified
Initiated
2023-09-12
Posted
2023-10-12
Terminated
not on file
Reason
Certain vials of the BD BACTEC" Plus Aerobic/F Culture Vials have a labeling issue where duplicate barcode sequence numbers have been identified on more than one vial.
Root cause
Under Investigation by firm
Action
An URGENT: Medical Device Product Correction notification letter dated 9/12/23 was sent to customers. Please Take the Following Actions: 1. Immediately inspect your inventory for the specific catalog and lot numbers listed above to determine if the BD BACTEC" Plus Aerobic/F Culture Vials in your possession are impacted. 2. Ensure the contents of this Product Advisory are read and understood. 3. Share and post this recall letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Users should verify that the accession number on the BD BACTEC" Instrument loading screen matches the vial's accession number while scanning and loading onto the instrument. If an error message or other system action occurs when loading vials into the BD BACTEC" Instrument that indicates a potential issue with duplicate sequence numbers, it is recommended to follow the instructions within the message and to view the table of System Alerts located in 7 Troubleshooting of the BD BACTEC" FX Instrument User s Manual before taking any further action in vial processing. The URL link to access the BD BACTEC" FX Instrument User s Manual is noted below. There are no additional follow-up activities recommended for patient samples processed. https://eifu.bd.com/en/search/search-results?term=441385&cats=9444_9445,9441_9443,9441_9442,1185_1206,1185 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 Distributors Please Ta
Quantity
781,600 vials
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 038290HXRPGWNNMN; Lot Numbers: 3062843 3062846 3062849 3062847 3067489 3067488
510(k) numbers
["K113558"]
PMA numbers
not on file
Product code
MDB
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register