Skip to the content

FDA Medical Device Recalls (openFDA)

Uoc Usa Inc: United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025

Recall number
Z-0110-2022
Recalling firm
Uoc Usa Inc
Product
United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025
Status
Terminated
Initiated
2021-08-27
Posted
not on file
Terminated
2023-01-20
Reason
Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.
Root cause
Device Design
Action
On 08/27/2021, the Recalling Firm communicated via email a letter to customers/US surgeons informing them that they are initiating a product recall limited to its 25 mm segment from its knee system due to an out of specification condition that has the potential to pose a risk to patients. Customers are instructed to return unused products. For questions, contact the Director of Customer Service at 949-328-3366 or the Vice President of Marketing at 949-501-8575. On 09/16/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" notification letter to additional customers, who have not used or received any of the affected product, just to make them aware of the recall. The Urgent Medical Device Recall notification identified the reason for the recall, the potential risks associate with the use of the affected product, and instructed these customers that to notify the recalling firm is they have any of the affected products. The Recalling Firm can be contacted via email harsh.gupta@unitedorthopedic.com or via phone at 949-501-8575. On 09/24/2021, the Recalling Firm communicated an updated "URGENT: MEDICAL DEVICE RECALL" Notification Letter informing customers that had received the affected recalled devices, of the recall action and the potential hazards associated with the affected products. In addition, this updated notification was also informing their customers, if they have any patients with the affected product implanted, the Recalling Firm's investigation did not reveal any reason to suggest that there is any imminent danger to the patient since the noted failure was primarily due to a fall from height. It is up to the medical discretion of the treating physicians regarding whether or not to notify the patient. It is also up to the medical discretion of the treating physician whether or not to consider a revision surgery to replace the affected segment. Any questions, contact: -Director of Customer Service at 949-328-3366, or -Vice President of
Quantity
3 devices
Distribution
U.S. Nationwide: CA, CO, and NV No O.U.S.
Lot, serial or model codes
Catalog Number: 2915-1025 UDI Code: 04719872148197 Lot Code: 19E272H
510(k) numbers
["K190100"]
PMA numbers
not on file
Product code
KRO
Firm FEI
3009760038
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28