FDA Medical Device Recalls (openFDA)
Ekos Corp: The Lysus Infusion System - Catheter Interface Cable. The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System consists of three main components, a single use Lysus Infusion Catheter System which is comprised of a Drug Delivery Catheter (DDC) and removable Ultrasound Core (USC), and a reusable EKOS PT-3 Control System. During use, fluids are delivered through the side-holes of the DDC. The USC is placed through the central lumen of the DDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS PT-3 Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Catheter Interface Cable with Pod connects the EKOS PT-3 Control Unit to the Lysus Infusion Catheter System.
- Recall number
- Z-0110-06
- Recalling firm
- Ekos Corp
- Product
- The Lysus Infusion System - Catheter Interface Cable. The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System consists of three main components, a single use Lysus Infusion Catheter System which is comprised of a Drug Delivery Catheter (DDC) and removable Ultrasound Core (USC), and a reusable EKOS PT-3 Control System. During use, fluids are delivered through the side-holes of the DDC. The USC is placed through the central lumen of the DDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS PT-3 Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Catheter Interface Cable with Pod connects the EKOS PT-3 Control Unit to the Lysus Infusion Catheter System.
- Status
- Terminated
- Initiated
- 2004-09-21
- Posted
- 2005-11-01
- Terminated
- 2005-11-08
- Reason
- The 100V board in the catheter interface cable pod failed during use. The PT-3 control unit would not recognize the catheter interface cable.
- Root cause
- Other
- Action
- On 9/21/04 the firm began visiting customers and replacing the catheter interface cables.
- Quantity
- 3
- Distribution
- Units were shipped to hospitals, one each in WA, AZ, and OH.
- Lot, serial or model codes
- Serial numbers: 10098-012, 10098-011, 10098-055 Part number: 3929-001 Catalog number: 700-10301
- 510(k) numbers
- ["K042456"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 3001627457
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28