Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott: TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter

Recall number
Z-0109-2026
Recalling firm
Abbott
Product
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter
Status
Open, Classified
Initiated
2025-09-10
Posted
2025-10-10
Terminated
not on file
Reason
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
Root cause
Package design/selection
Action
Abbott issued a Medical Device Advisory to its consignees on 09/10/2025 by letter delivered via Federal Express. The notice explained the issue, impact and associated risk to the patient and requested the following: "Action Requested/Recommendations/Next Steps . Continue following the product IFU by inspecting the packaging and catheter for damage and/or kinking prior to use. . Review the guidance in Appendix A information on proper removal technique of the TactiFlex" Ablation Catheter, Sensor Enabled" from the packaging tray, ensuring to inspect the electrodes and catheter tip carefully for integrity and overall condition prior to use. . Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal from the packaging. The deployment of the new packaging has begun. Please complete the provided Acknowledgement Form with the assistance of your Abbott representative. This form ensures you are aware of the advisory notice and have been given an opportunity to ask questions. Please forward this notice to anyone in your organization who may need to be notified and maintain this notice for your records to ensure effectiveness of this communication. Should you have questions about this issue, please contact your local Abbott representative or Abbott Support at 1-855-478-5833 (U.S), 8:30 a.m. - 5:30 p.m. Central Standard Time, Monday thru Friday."
Quantity
407 units
Distribution
Worldwide
Lot, serial or model codes
GTIN 05415067034656, Batch Numbers: 8894231, 8915981, 9029640, 9255006, 10001631, 10053399, 10103653, 10120408, 10151069, 10194386, 10194388, 10217380, 10389653, 10429584, 10440518, 10513213, 10513225, 10513232, 10513235, 10551326, 10562041, 10562047, 10562049, 10566077, 10597231, 10597242, 10597249, 10597260, 10613572, 10613581, 10613583, 10613585, 10632079, 10632621, 10632626, 10632634, 10646249, 10673011, 10675350, 10676858, 10677749, 10680007, 10757210, 10785266, 10798695, 10798697, 10816723, 10817847, 10817849, 10850906, 10850907, 10858157, 10858160, 10860077, 10871703, 10885761, 10885921, 10889919, 10892126, 10896652, 10898353, 10899933, 10900171, 10903297, 10903302, 10909596, 10910449, 10910454, 10913639, 10914972, 10917505, 10918937, 10922625, 10922639, 10928017, 10938688, 10939318, 10939319, 10939322, 10950790, 10954012, 10954308, 10954309, 10956216, 10956223, 10957148, 10960327, 10960343, 10961378, 10962731, 10956836, 10958929, 10984316, 10984788, 10985584, 10987173, 11076153, 10984516, 10985598, 10962737.
510(k) numbers
not on file
PMA numbers
["P220013"]
Product code
OAE
Firm FEI
3005334138
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register