Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE HEALTHCARE, MSK 1.5T EXTREME MR SCANNER, MODEL AA5000. The MSK Extreme TM MR Scanner is intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand.

Recall number
Z-0107-2012
Recalling firm
GE Healthcare, LLC
Product
GE HEALTHCARE, MSK 1.5T EXTREME MR SCANNER, MODEL AA5000. The MSK Extreme TM MR Scanner is intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand.
Status
Terminated
Initiated
2011-08-29
Posted
2011-10-26
Terminated
2012-05-29
Reason
GE Healthcare has recently become aware that an exposed high voltage gradient wire internal to the MSK 1.5T Extreme system equipment may cause electrical shock, which may impact service personnel safety. This issue does not impact patient or operator safety.
Root cause
Nonconforming Material/Component
Action
GE Healthcare LLC sent a "Urgent Medical Device Correction letter" dated August 29, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to please ensure that all potential service personnel at your facility are made aware of this safety notification and the recommended actions. Contact your local service representative if you have any questions concerning this notification. For further questions please call ( 262 ) 513-4122.
Quantity
52
Distribution
Worldwide Distribution -- USA (nationwide) including the states of AL, CA, FL, ID, NY, NH, NJ, NV, OH, PA, TX, VA and WI., and countries of UNITED ARAB EMIRATES, UKRAINE, UNITED KINGDOM, SWITZERLAND
Lot, serial or model codes
06012008-001 12182008-001 12162008-001  03132009-001  03242009-001 05152009-001  07142009-001  08082009-001 09172009-001 12072009-001 10252009-001 01082010-001 012720100-001 03072010-001 03252010-001 04012010-001 05262010-001 06082010-001 06182010-001 09142010-001 11292010-001 10182010-001 12052010-001 08122009-001 09222009-001 02242010-001 10082009-001 02112010-001 10222010-001 08132009-001 12222010-001  06192009-001 10312010-001 10142009-001 09222010-001 10012009-001 06042009-001  06062009-001 09072009-001 01182010-001 03012010-001 07272010-001 09132010-001 10082010-001 11162010-001 12152009-001 12242009-001 12142009-001 04302010-001 04082010-001 05072010-001 08152010-001
510(k) numbers
["K080048"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register