FDA Medical Device Recalls (openFDA)
Helena Laboratories, Inc.: "Free and Total Protein S ELISA Kit" Cat. No. 5292 For use in the quantitation of free and total Protein antigen in citrated human plasma Contains: 96 Protein S: Ag Microwells, 3 x 0.5mL ELISA Reference Plasma, 1x12 mL Protein S Conjugate Solution, 1 x 60mL Sample Diluent, 1 x 13mL Substrate, 1 x 15mL Stopping Solution, 1 x 30mL Phosphate Buffered Saline, 1 x 2mL Free Protein S Reagent FOR IN-VITRO DIAGNOSTIC USE LOT: 3-16- 5292 EXP: 2018-01-26 HELENA LABORATORIES BEAUMONT, TX Outer labeling of Substrate (brown bottle and cap): "Substrate 13 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26" Outer Labeling of Protein S Conjugate bottle (clear bottle and cap, red liquid): "Protein S Conjugate 12 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26"
- Recall number
- Z-0106-2019
- Recalling firm
- Helena Laboratories, Inc.
- Product
- "Free and Total Protein S ELISA Kit" Cat. No. 5292 For use in the quantitation of free and total Protein antigen in citrated human plasma Contains: 96 Protein S: Ag Microwells, 3 x 0.5mL ELISA Reference Plasma, 1x12 mL Protein S Conjugate Solution, 1 x 60mL Sample Diluent, 1 x 13mL Substrate, 1 x 15mL Stopping Solution, 1 x 30mL Phosphate Buffered Saline, 1 x 2mL Free Protein S Reagent FOR IN-VITRO DIAGNOSTIC USE LOT: 3-16- 5292 EXP: 2018-01-26 HELENA LABORATORIES BEAUMONT, TX Outer labeling of Substrate (brown bottle and cap): "Substrate 13 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26" Outer Labeling of Protein S Conjugate bottle (clear bottle and cap, red liquid): "Protein S Conjugate 12 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26"
- Status
- Terminated
- Initiated
- 2017-04-04
- Posted
- not on file
- Terminated
- 2020-05-14
- Reason
- Internal components of the Protein S ELISA Antigen Kit (Lot: 3-16-5292) were mislabeled. The Protein S Conjugate Solution is labeled as Substrate, and the Substrate is labeled as Protein S conjugate.
- Root cause
- Labeling mix-ups
- Action
- On 4/4/17, an email notification was initially sent to the firm's OUS distributor that received 3 kits. On 4/10/17, the customer replied with a photograph to confirm their kits were affected. A Withdrawal notice was approved internally, and formally issued by the firm to a single distributor via e-mail on 4/17/17. An authorized confirmation of destruction was sent from the distributor to the firm via e-mail on 4/18/17. The firm's public contact may be reached at (409) 842-3714 ext. 1177.
- Quantity
- 3 kits
- Distribution
- International distribution to country of: Jordan
- Lot, serial or model codes
- 3-16-5292; UDI: Primary DI - M52552920
- 510(k) numbers
- ["K972482"]
- PMA numbers
- not on file
- Product code
- GGP
- Firm FEI
- 1618982
- Firm city
- Beaumont
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28