Varian Medical Systems Oncology Systems: Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian Medical Systems Inc., Palo Alto, CA
- Recall number
- Z-0106-2011
- Recalling firm
- Varian Medical Systems Oncology Systems
- Product
- Varian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian Medical Systems Inc., Palo Alto, CA
- Status
- Terminated
- Initiated
- 2010-05-24
- Posted
- 2010-10-20
- Terminated
- 2011-07-20
- Reason
- Due to a software anomaly, guidance-based couch shift values may not be applied as expected when the operator selects "Apply Shift" and presses Motion Enable buttons. This can result in the actual patient position differing from that position indicated by the imaging application affecting poistion accuracy and precision.
- Root cause
- Software design
- Action
- Varian Medical Systems issued an Urgent Medical Device Correction letter date dated May 24, 2010 to users identifying the affected device, the issue prompting the correction, and the actions to be taken by the user. Customers were asked to set couch tolerance values to small numbers, and for radiosurgical treatment tolerance values to be set to zero. A Varian representative will contact customers to schedule installation of a software version correcting the issue. US Customers can contact Varian at 1-888-827-4265,. European customers can contact Varian at +41 41 749 8844.
- Quantity
- not on file
- Distribution
- Worldwide Distribution: Two consignees, one in the US and one in Switzerland.
- Lot, serial or model codes
- H191001, H191003
- 510(k) numbers
- ["K092871"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 2916710
- Firm city
- Palo Alto
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28