Skip to the content

FDA Medical Device Recalls (openFDA)

Cti Pet Systems Inc: Siemens Models: BIOGRAPH, BIOGRAPH LSO and CTI Models: REVEAL HD, REVEAL RT

Recall number
Z-0106-04
Recalling firm
Cti Pet Systems Inc
Product
Siemens Models: BIOGRAPH, BIOGRAPH LSO and CTI Models: REVEAL HD, REVEAL RT
Status
Terminated
Initiated
2003-10-02
Posted
2003-11-13
Terminated
2006-02-07
Reason
CPS has found that in some instances, when the CT E-Stop is activated, a malfunction may occur where the PHS controller may not shut down and stop bed motion.
Root cause
Other
Action
Written notification was sent out beginning 10/02/2003 to distributors and end users to advise them of the problem.
Quantity
73
Distribution
Nationwide International: Japan, Germany, Brazil, Spain, China, Russia, Austrailia, Hong Kong, Singapore, Poland France, Korea, Greece, UK and Austria
Lot, serial or model codes
All CPS manufactured PET/CT systems distributed between October 31, 2000 and June 24, 2003.
510(k) numbers
["K013504","K002715"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1034973
Firm city
Knoxville
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register