FDA Medical Device Recalls (openFDA)
Fresenius Medical Care Holdings, Inc.: LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
- Recall number
- Z-0105-2011
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Product
- LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
- Status
- Terminated
- Initiated
- 2010-09-03
- Posted
- 2010-10-20
- Terminated
- 2023-02-16
- Reason
- Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.
- Root cause
- Nonconforming Material/Component
- Action
- Fresenius Medical Care issued an Urgent FMCNA LIBERTY Cycler Set Recall letter on September 3, 2010 identifying the specific affected part numbers and product lots and requesting that customers identify and discontinue the use of the affected lots immediately. Customers were requested to return a receipt of notification. Customers may arrange return unused product to FMCNA by contacting customer service at 1-800-323-5188.
- Quantity
- 5799 units
- Distribution
- Nationwide Distribution: Throughout USA.
- Lot, serial or model codes
- Lot Number Expiration Date 09ER08101 05/31/2012 09HR08002 06/30/2012 09HR08003 06/30/2012 09JR08057 07/31/2012 09JR08058 07/31/2012 09JR08059 07/31/2012 09KR08002 08/31/2012 09KR08003 08/31/2012 09KR08004 08/31/2012 09KR08005 08/31/2012 09LR08004 09/30/2012 09LR08005 09/30/2012 09LR08006 09/30/2012 09LR08007 09/30/2012 09LR08008 09/30/2012 09LR08010 09/30/2012 09LR08105 09/30/2012
- 510(k) numbers
- ["K043363"]
- PMA numbers
- not on file
- Product code
- FKX
- Firm FEI
- 3001451489
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28