Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.

Recall number
Z-0105-2011
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
LIBERTY Automated Peritoneal Dialysis System lntegrated Cycler Set with single stay safe patient connector and extended drain line PIN: 050-87216. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
Status
Terminated
Initiated
2010-09-03
Posted
2010-10-20
Terminated
2023-02-16
Reason
Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.
Root cause
Nonconforming Material/Component
Action
Fresenius Medical Care issued an Urgent FMCNA LIBERTY Cycler Set Recall letter on September 3, 2010 identifying the specific affected part numbers and product lots and requesting that customers identify and discontinue the use of the affected lots immediately. Customers were requested to return a receipt of notification. Customers may arrange return unused product to FMCNA by contacting customer service at 1-800-323-5188.
Quantity
5799 units
Distribution
Nationwide Distribution: Throughout USA.
Lot, serial or model codes
Lot Number Expiration Date 09ER08101 05/31/2012 09HR08002 06/30/2012 09HR08003 06/30/2012 09JR08057 07/31/2012 09JR08058 07/31/2012 09JR08059 07/31/2012 09KR08002 08/31/2012 09KR08003 08/31/2012 09KR08004 08/31/2012 09KR08005 08/31/2012 09LR08004 09/30/2012 09LR08005 09/30/2012 09LR08006 09/30/2012 09LR08007 09/30/2012 09LR08008 09/30/2012 09LR08010 09/30/2012 09LR08105 09/30/2012
510(k) numbers
["K043363"]
PMA numbers
not on file
Product code
FKX
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register