Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems LLC: Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System

Recall number
Z-0105-06
Recalling firm
General Electric Med Systems LLC
Product
Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System
Status
Terminated
Initiated
2005-09-27
Posted
2005-10-28
Terminated
2006-12-26
Reason
Cracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.
Root cause
Other
Action
GE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.
Quantity
140 (80 US and 60 OUS)
Distribution
Within the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain, Sweden and United Kingdom
Lot, serial or model codes
Cameras affected by this safety notice are labeled with the SOPHA MEDICAL logo. STATIF DST and STATIF DST-XL model 100, serial numbers 001-264. The label is located on the inner side of the gantry cover. All affected cameras were manufactured prior to 1997 by Sopha Medial (SMV), which was subsequently acquired by GE Healthcare in 2000.
510(k) numbers
["K942837"]
PMA numbers
not on file
Product code
KPS
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register