Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay

Recall number
Z-0104-2023
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay
Status
Open, Classified
Initiated
2022-09-19
Posted
2022-10-20
Terminated
not on file
Reason
Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.
Root cause
Under Investigation by firm
Action
On September 19, 2022, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers should take the following actions: -Please review this letter with your Medical Director. -Discontinue use of and discard the affected product. Siemens is temporarily discontinuing production and distribution of the Atellica CH lithium reagents until the issue is resolved. In the interim, Siemens is recommending that customers transition to an alternate lithium assay.
Quantity
1342 US; 2266 OUS
Distribution
Worldwide distribution - US Nationwide.
Lot, serial or model codes
UDI-DI: 00630414287935 All lots impacted: Lots 110238 exp 10/18/22 110302 exp 12/13/22 120017 exp 4/18/23
510(k) numbers
["K050374"]
PMA numbers
not on file
Product code
NDW
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register