Roche Diagnostics Operations, Inc.: Albumin Gen.2, Catalog 05166861190

Recall number
Z-0104-2021
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Albumin Gen.2, Catalog 05166861190
Status
Terminated
Initiated
2020-08-27
Posted
not on file
Terminated
2024-05-22
Reason
Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.
Root cause
Under Investigation by firm
Action
On August 27, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that this was an expansion of the previous notification sent on September 11, 2019, as new products and lots had been added to the scope of the recall. Customers were asked to do the following: " Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. " Refer to the Product Replacement section of the letter for product replacement information. " Complete all sections of the enclosed fax back form and fax or email it according to the instructions on the form even if you are not requesting product replacement. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. Roche will replace any of the recalled product remaining in your inventory. To receive replacement product, please complete the response form enclosed with the firm's mailing and fax or email it according to the instructions on the form. Please contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about this recall.
Quantity
11963
Distribution
Domestic distribution nationwide.
Lot, serial or model codes
lots 43031001, 43718901
510(k) numbers
["K033009"]
PMA numbers
not on file
Product code
CIX
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register