Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems, Inc.: Philips IntelliVue Monitors with software revisions J.21.03, J.21.19. Model Product MP5 M8105A; MP5SC M8105AS; MP5T M8105AT; MP5 Upgrade M8105AU ; MP5SC Upgrade 866327 The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.

Recall number
Z-0104-2015
Recalling firm
Philips Medical Systems, Inc.
Product
Philips IntelliVue Monitors with software revisions J.21.03, J.21.19. Model Product MP5 M8105A; MP5SC M8105AS; MP5T M8105AT; MP5 Upgrade M8105AU ; MP5SC Upgrade 866327 The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Status
Terminated
Initiated
2014-10-03
Posted
2014-10-21
Terminated
2017-01-31
Reason
Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped
Root cause
Software Design Change
Action
Philips Healthcare issued on 10/3/14 the Urgent Medical Device Correction notification/Field Safety Notice. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a software upgrade free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for service. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: During the interim period until the SW is upgraded please make sure, if the NBP repetition interval needs to be increased (e.g. from 5 to 15 minutes), to restart the automatic NBP measurement. Please review this information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.
Quantity
7587 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN KAZAKHSTAN NETHERLANDS NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN POLAND ROMANIA RUSSIA SINGAPORE SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE and UNITED KINGDOM.
Lot, serial or model codes
software revisions J.21.03, J.21.19
510(k) numbers
["K122439","K131913"]
PMA numbers
not on file
Product code
MHX
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register