FDA Medical Device Recalls (openFDA)
Fresenius Medical Care Holdings, Inc.: LIBERTY Automated Peritoneal Dialysis System Line Cycler Set with stay safe connectors (Single Patient Connector) PIN: 050-87215. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
- Recall number
- Z-0104-2011
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Product
- LIBERTY Automated Peritoneal Dialysis System Line Cycler Set with stay safe connectors (Single Patient Connector) PIN: 050-87215. manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, Mexico.
- Status
- Terminated
- Initiated
- 2010-09-03
- Posted
- 2010-10-20
- Terminated
- 2023-02-16
- Reason
- Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.
- Root cause
- Nonconforming Material/Component
- Action
- Fresenius Medical Care issued an Urgent FMCNA LIBERTY Cycler Set Recall letter on September 3, 2010 identifying the specific affected part numbers and product lots and requesting that customers identify and discontinue the use of the affected lots immediately. Customers were requested to return a receipt of notification. Customers may arrange return unused product to FMCNA by contacting customer service at 1-800-323-5188.
- Quantity
- 1524 units
- Distribution
- Nationwide Distribution: Throughout USA.
- Lot, serial or model codes
- Lot Number Expiration Date 09CR08086 03/31/2012 09DR08038 04/30/2012 09JR08092 07/31/2012 09JR08093 07/31/2012 09KR08061 08/31/2012
- 510(k) numbers
- ["K043363"]
- PMA numbers
- not on file
- Product code
- FKX
- Firm FEI
- 3001451489
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28